# Build the circuit netlist required for extraction

Describe the branch model and port-to-node topology that InductEx® fits, with consistent design units and a complete microstrip netlist.

## Compatibility

8 beta/preview

Published guidance with release-applicability limits. Confirm package applicability before following this guide.

**KB-EDITOR-001 · supported path**: The checked extraction netlist accepts .end while its editor reports this specific unsupported-directive warning. Preserve the complete input and check extraction independently; do not suppress unrelated warnings.

**KB-GUI-006 · verified workaround**: Save work and close other saved tabs before exit when the intended Studio project must return as active. The checked single-tab restart does not repair multi-tab selection.

**KB-GUI-008 · verified workaround**: After a source edit, save the intended inputs, choose Run again and verify the new result. An earlier Completed label may still refer to the previous generation.

[Download Complete inline MQS microstrip inputs (ZIP)](/_functions/kbExample?name=inductex-inline-microstrip.zip)

Includes: strip.ixi, process.ldf, strip.cir, README.txt.

## Before starting

This is an InductEx® extraction topology. Its P/J/B/I port identifiers describe excitation connections; do not paste a complete JoSIM device/transient deck into this file. Download the shared complete example for the matching process and layout.

## Steps

1. Create or open strip.cir as an InductEx® Netlist in Studio. Start any title with *; a bare SPICE title line is not the InductEx® title convention.
2. Define each extracted branch with its element name and positive/negative node names. In this example L1 connects node 1 to node 2.
3. Declare P1 1 0 and P2 2 0. Match P1/P2 to the layout labels and connect node 0 to the intended common negative terminal, here M0.
4. Enter design values in the units specified by UnitL/UnitR in the process. Save strip.cir and confirm strip.ixi references it before running the project.
5. Inspect current run status, recovered terminal mapping, requested branch names, units and fit/design diagnostics. Numerical fit quality is separate from whether the topology describes the real circuit.

![The native extraction-netlist editor shows L1 between nodes 1 and 2 and ports P1 and P2 returning to node 0, with an editor warning on .end.](https://static.wixstatic.com/media/31de30_8cf631d4b9a24e898a38dc41dd4fcb79~mv2.png)

The complete L1/P1/P2 extraction netlist in the actual editor. The candidate’s .end warning is a reproduced diagnostic mismatch, explained below.

## Complete strip.cir

```text
* Extraction topology, design inductance in henries because UnitL=1
L1 1 2 3.9e-12
P1 1 0
P2 2 0
.end
```

## Read the topology and design value

L1 1 2 3.9e-12 defines the branch model between the two strip ends; P1 and P2 return those ends to node 0 through the negative terminal. Node names describe abstract circuit connectivity and need not encode geometric positions.

With UnitL = 1 the design value is 3.9e-12 H. Design values are optional in the maintained format; omitting them removes that design comparison rather than defining a zero-inductance target.

For mutual terms K/M refer to inductor names rather than node names. Choose a topology that the available excitations can identify. Extra branches do not create independent information, and a negative fitted value is not automatically an invalid extraction. Use the dedicated modelling guides for fit interpretation.

## Check the result

The checked 100 × 10 µm example gives L1 = 3.896479063e-12 H (3.896479063 pH), approximately 0.09028% below its 3.9 pH design value. Its mesh bytes and extracted value match the maintained GDS microstrip. This comparison is specific to these exact inputs and runtime.

The one-branch teaching example is not evidence that an arbitrary multi-branch circuit is identifiable or calibrated. Inspect fit rank and residuals along with geometry and excitation coverage.

The detailed current CLI output/elements.txt contains the value quoted above in the configured UnitL units. The human-readable summary rounds this example to 3.89648 pH. Match units and printed precision before treating different displays as a numerical change.

## If the problem remains

An incomplete line such as L1 1 is rejected by the production editor analysis. Restore both terminal nodes before checking the design value. A P2/P9 mismatch between layout and netlist was rejected by the CLI and recovered after restoring the intended name.

NETLIST_UNSUPPORTED_DIRECTIVE on .end is a reproduced editor/CLI mismatch in this candidate (KB-EDITOR-001). The checked 8.0.1 CLI accepts this terminator and extracts the circuit. Inspect other warnings and the actual run outcome separately; an editor warning is not itself an extraction result.

For JoSIM handoff, use the current IXI back-annotation request and a separate simulation template. The historical netlist BACK-ANNOTATE location has a version-specific trap; follow the dedicated checked handoff guide.

## Actual native result check

The native workflow opened and edited this complete circuit, ran the normal engine and recovered after correcting a P2/P9 mismatch in the layout labels. Match the current result units to UnitL; the netlist’s optional design value does not set an accuracy tolerance.

![Current native raw output gives the 100 micrometre strip L1 as 3.89648e-12 H against the 3.9e-12 H design value.](https://static.wixstatic.com/media/31de30_fd644d62c24744dda9ce716d07b93b42~mv2.png)

Current native L1 result for this one-branch topology, with the original design and result values in H.

## Sources and verification details

Recorded inputs/builds only. Complete walkthrough and final-package acceptance remain separate.

- VAL-020: Windows catalog InductEx 8.0.1 (2 November 2026) and its bundled TTH 1.1.50; not final-package acceptance. Complete inline MQS strip, byte-identical mesh and identical inductance versus the maintained GDS example, length/display-unit changes, one mesh refinement and missing-process/port/layer rejection and recovery Sixteen bounded CLI checks. One refinement observation is not a convergence threshold. Design agreement is not physical calibration or solver/fit residual acceptance. Native GUI authoring, cancellation/retry, restart and final distributed-package checks remain pending.
- VAL-021: Frozen current Windows Studio editor/nominal-process analysis source, 3 November 2026. Seventeen production-parser/analysis checks: exact input syntax, three-layer nominal stack/units, malformed input and restoration, plus focused reproduction of TERM/TEXT annotation alias warnings This route is editor analysis only. It does not construct the Core model, run an extraction, verify physical calibration or complete a native GUI walkthrough. The accepted .end directive still receives an editor warning (KB-EDITOR-001); annotation alias warnings remain KB-GUI-005.
- VAL-023: Frozen Windows Studio source candidate with normal Core 8.0.1 dispatch, 3 November 2026; not a final distributed package. Normal production source edits/Undo/Redo/Save, licensed baseline/length/unit runs, port failure and correction, Stop/retry and a separate-process single-tab restart/rerun. File/directory selection used predetermined paths at the production dialog boundary Eighteen independent checks cover these stated routes. The prior Completed label after a saved source edit remains KB-GUI-008; stale-state acceptance is not passed. A Cancelled receipt retained an L1 row and is not accepted as a completed solve. Physical pickers, complete GUI article, final distributed package, other platforms, physical accuracy and mesh convergence remain pending. KB-GUI-006 single-tab restart workaround remains separate.

- studio-extraction-netlist: Current Windows Studio source, native authoring illustration, 3 November 2026. Authoring illustration only; no GUI extraction or complete walkthrough is shown. CLI/editor results are recorded separately. Annotation alias and .end warnings remain under review.
- studio-inline-current-inductance: Frozen Windows Studio source candidate, actual native workflow, 3 November 2026. This image shows the stated recorded control/result only. The retained Completed-after-edit observation remains open; a Cancelled receipt is not accepted numerical success. Complete GUI article and final-package checks remain pending.

- inductex:docs/manual/ix_netlist.tex: Title/comments, elements, design values and ports; BACK-ANNOTATE limits remain separate
- inductex:docs/manual/ix_ports.tex: Layout terminal polarity matched to netlist nodes
- inductex:docs/manual/studio_projects.tex: Create/open a netlist and run its owning IXI project
- validation:inductex-inputs-windows-20261103: VAL-020 licensed CLI and VAL-021 separate editor-analysis checks; exact scope retained
- capture:studio-inputs-20261103T163127: Native authoring illustration only; exact reviewed input and image bytes
- validation:studio-inputs-native-windows-20261103: VAL-023; actual native edits, numerical runs, port recovery, Stop/retry and separate-process single-tab restart; provenance limitation retained

- The five cited maintained references still match their reviewed inventory hashes; existing source baselines were preserved.
- VAL-023 adds eighteen independent native checks. The saved-source completion label is an open provenance observation (KB-GUI-008); the actual Cancelled verdict is kept distinct from retained numerical rows. Original input bytes were restored.
- VAL-020 CLI and VAL-021 editor analysis remain separate from the later normal native workflow. Complete physical pickers, article/release acceptance and numerical qualification remain pending.

## Related guides

- [Create parameterised IXI geometry without a GDS file](/knowledge-base/inductex-ixi-geometry)
- [Define and inspect an LDF fabrication process](/knowledge-base/inductex-process-build)
- [Define layout ports and match them to the circuit](/knowledge-base/inductex-port-design)