{"sections":[{"figures":[{"asset_id":"studio-saved-length120","caption":"Actual saved source contains strip_length = 120 before the next extraction."},{"asset_id":"studio-prior-completed-after-edit","caption":"At that point the actual Results status row still read Completed for the retained earlier calculation. This is an open provenance observation, not a passed stale-state check."}],"heading":"Recognise the retained result","paragraphs":["This is a reproduced observation on a frozen Windows Studio source candidate, not a claim about every released package. After a completed 100 µm strip run, changing strip_length to 120 in Raw IXI, using Undo/Redo and saving left the earlier Results status reading Completed.","That label alone does not prove that the current source has been calculated. KB-GUI-008 tracks how the last-run result and edited source should be distinguished. The full GUI article and final-package behavior remain unqualified."]},{"heading":"Steps","steps":["Save all intended inputs. Confirm the active IXI and its process, circuit and variable values; keep a record of the input revision with the result.","Use Check and run / Run full extraction. After a previous completion Basic labels the production action Run again rather than Run extraction.","Wait for the new task verdict. Open Results > Raw output and check the quantities and units for that run, along with any errors or warnings.","If the run fails or is cancelled, correct the cause or finish the intended edits and run again. Do not accept an old file, the earlier Completed label or an isolated numeric row as a new successful result."]},{"figures":[{"asset_id":"studio-inline-longer-inductance","caption":"The new native run uses the saved 120 µm input and produces the larger L1. This is the result to compare with that edited input."}],"heading":"Check the result","paragraphs":["For the complete teaching example, the original 100 µm strip gives 3.89648e-12 H at GUI printed precision. Run again after saving 120 µm gives 4.67906e-12 H, matching the independently checked CLI at the same printed precision. Reverting the original input values and rerunning recovers the baseline.","The 3.9 pH design target was deliberately left unchanged during this length demonstration. Its resulting design deviation does not measure solver accuracy. Review the design target and mesh convergence separately for the actual problem."]},{"heading":"Cancellation and restart","paragraphs":["The recorded Stop action reached Cancelled while its raw output still contained an L1 row. The verdict remains Cancelled; that receipt is not accepted as a completed solve. Normal retry completed successfully.","A separate process restored the corrected saved source with the single-tab workaround, but Results was initially empty. Retained solver files did not establish a new completion. Deliberately rerunning produced the baseline. KB-GUI-006 remains separate from this after-edit finding."]},{"heading":"If the problem remains","paragraphs":["Read the specific current error, verify saved paths and input revision, and use the linked missing-input, licensing or port guide as appropriate. If reporting the provenance issue, include the exact application/source version, actions and a minimal public reproducer; review any logs for private design or licence information before sharing."]}],"canonical_url":"https://www.sun-magnetics.com/knowledge-base/inductex-completed-after-edit","url":"/knowledge-base/inductex-completed-after-edit","known_issues":[{"disposition":"verified_workaround","id":"KB-GUI-006","title":"Normal multi-tab exit saves the last tab as active","public_notice":"Save work and close other saved tabs before exit when the intended Studio project must return as active. The checked single-tab restart does not repair multi-tab selection.","claim_scope":"Only the affected claim and recorded build/input; not whole-guide or final-release acceptance"},{"disposition":"verified_workaround","id":"KB-GUI-008","title":"A saved source edit leaves the prior Completed result visible","public_notice":"After a source edit, save the intended inputs, choose Run again and verify the new result. An earlier Completed label may still refer to the previous generation.","claim_scope":"Only the affected claim and recorded build/input; not whole-guide or final-release acceptance"}],"search_record":{"execution":true,"relatedProducts":[],"version_tags":["InductEx 8 preview"],"text":"Results still says Completed after I edit an IXI file The checked Studio candidate can retain the earlier Completed result after a saved source edit. Save the intended inputs, use Run again and check the new verdict and values. InductEx 8 beta/preview Troubleshooting Completed result after editing retained last run stale source Run again saved length 120 KB-GUI-008 source result identity Recognise the retained result This is a reproduced observation on a frozen Windows Studio source candidate, not a claim about every released package. After a completed 100 µm strip run, changing strip_length to 120 in Raw IXI, using Undo/Redo and saving left the earlier Results status reading Completed. That label alone does not prove that the current source has been calculated. KB-GUI-008 tracks how the last-run result and edited source should be distinguished. The full GUI article and final-package behavior remain unqualified.  Actual saved source contains strip_length = 120 before the next extraction. Actual saved Raw IXI source shows strip_length set to 120, alongside the same process and netlist references. At that point the actual Results status row still read Completed for the retained earlier calculation. This is an open provenance observation, not a passed stale-state check. The actual Results status row still reads Completed after saving a changed strip length, before Run again; this is the retained earlier result. Steps Save all intended inputs. Confirm the active IXI and its process, circuit and variable values; keep a record of the input revision with the result. Use Check and run / Run full extraction. After a previous completion Basic labels the production action Run again rather than Run extraction. Wait for the new task verdict. Open Results > Raw output and check the quantities and units for that run, along with any errors or warnings. If the run fails or is cancelled, correct the cause or finish the intended edits and run again. Do not accept an old file, the earlier Completed label or an isolated numeric row as a new successful result.  Check the result For the complete teaching example, the original 100 µm strip gives 3.89648e-12 H at GUI printed precision. Run again after saving 120 µm gives 4.67906e-12 H, matching the independently checked CLI at the same printed precision. Reverting the original input values and rerunning recovers the baseline. The 3.9 pH design target was deliberately left unchanged during this length demonstration. Its resulting design deviation does not measure solver accuracy. Review the design target and mesh convergence separately for the actual problem.  The new native run uses the saved 120 µm input and produces the larger L1. This is the result to compare with that edited input. After saving length 120 and using Run again, current native raw output gives L1 as 4.67906e-12 H. The unchanged design target is still 3.9e-12 H. Cancellation and restart The recorded Stop action reached Cancelled while its raw output still contained an L1 row. The verdict remains Cancelled; that receipt is not accepted as a completed solve. Normal retry completed successfully. A separate process restored the corrected saved source with the single-tab workaround, but Results was initially empty. Retained solver files did not establish a new completion. Deliberately rerunning produced the baseline. KB-GUI-006 remains separate from this after-edit finding.  If the problem remains Read the specific current error, verify saved paths and input revision, and use the linked missing-input, licensing or port guide as appropriate. If reporting the provenance issue, include the exact application/source version, actions and a minimal public reproducer; review any logs for private design or licence information before sharing. ","id":"inductex-completed-after-edit","keywords":"Completed result after editing retained last run stale source Run again saved length 120 KB-GUI-008 source result identity","kind":"Troubleshooting","title":"Results still says Completed after I edit an IXI file","product":"InductEx"},"version":"8 beta/preview","id":"inductex-completed-after-edit","related":["/knowledge-base/inductex-ixi-geometry","/knowledge-base/inductex-read-results","/knowledge-base/inductex-run-unavailable","/knowledge-base/inductex-missing-input"],"downloads":[{"url":"/_functions/kbExample?name=inductex-inline-microstrip.zip","scope":"Exact checked input files plus instructions; final-release/platform applicability remains pending.","files":["strip.ixi","process.ldf","strip.cir","README.txt"],"id":"inductex-inline-microstrip","requires":["Matching licensed InductEx 8/TTH runtime","New writable directory; keep strip.ixi, process.ldf and strip.cir together","Review the stated Windows 8.0.1 scope; complete GUI/final-package acceptance remains pending"],"title":"Complete inline MQS microstrip inputs"}],"publication_status":"published","numerical_output":null,"provenance":{"execution_checks":[{"routes":["gui-native-scenario"],"scope":"Normal production source edits/Undo/Redo/Save, licensed baseline/length/unit runs, port failure and correction, Stop/retry and a separate-process single-tab restart/rerun. File/directory selection used predetermined paths at the production dialog boundary","result":"passed","articles":["inductex-ixi-geometry","inductex-process-build","inductex-netlist-build","inductex-port-design","inductex-read-results","inductex-completed-after-edit"],"prior_result_after_edit":{"status_after_saved_length_edit":"Completed","stale_state_accepted":false,"issue":"KB-GUI-008","source_has_length_120":true,"source_is_saved":true},"purpose":"studio-input-edit-run-recovery-cancel-restart","version":"Frozen Windows Studio source candidate with normal Core 8.0.1 dispatch, 3 November 2026; not a final distributed package","id":"VAL-023","cancellation":{"accepted_as_numerical_success":false,"normal_retry_passed":true,"raw_output_contains_inductance_row":true,"verdict":"Cancelled"},"records":["native-inputs","native-inputs-restart","independent-input-review","exact restored input hashes"],"limit":"Eighteen independent checks cover these stated routes. The prior Completed label after a saved source edit remains KB-GUI-008; stale-state acceptance is not passed. A Cancelled receipt retained an L1 row and is not accepted as a completed solve. Physical pickers, complete GUI article, final distributed package, other platforms, physical accuracy and mesh convergence remain pending. KB-GUI-006 single-tab restart workaround remains separate.","observed":{"port-recovery":{"L1_H":3.89648e-12,"L1_display":3.89648e-12,"design_display":3.9e-12,"unit":"H"},"baseline":{"L1_H":3.89648e-12,"L1_display":3.89648e-12,"design_display":3.9e-12,"unit":"H"},"restarted":{"L1_H":3.89648e-12,"L1_display":3.89648e-12,"design_display":3.9e-12,"unit":"H"},"longer":{"L1_H":4.67906e-12,"L1_display":4.67906e-12,"design_display":3.9e-12,"unit":"H"},"after-cancel":{"L1_H":3.89648e-12,"L1_display":3.89648e-12,"design_display":3.9e-12,"unit":"H"},"picohenry":{"L1_H":3.8964799999999996e-12,"L1_display":3.89648,"design_display":3.9,"unit":"pH"}}}],"figure_evidence":[{"evidence_limit":"This image shows the stated recorded control/result only. The retained Completed-after-edit observation remains open; a Cancelled receipt is not accepted numerical success. Complete GUI article and final-package checks remain pending.","id":"studio-saved-length120","reference_label":"Frozen Windows Studio source candidate, actual native workflow, 3 November 2026"},{"evidence_limit":"This image shows the stated recorded control/result only. The retained Completed-after-edit observation remains open; a Cancelled receipt is not accepted numerical success. Complete GUI article and final-package checks remain pending.","id":"studio-prior-completed-after-edit","reference_label":"Frozen Windows Studio source candidate, actual native workflow, 3 November 2026"},{"evidence_limit":"This image shows the stated recorded control/result only. The retained Completed-after-edit observation remains open; a Cancelled receipt is not accepted numerical success. Complete GUI article and final-package checks remain pending.","id":"studio-inline-longer-inductance","reference_label":"Frozen Windows Studio source candidate, actual native workflow, 3 November 2026"}],"review_notes":["VAL-023 records eighteen independently checked native actions/outcomes with original teaching input bytes restored. The actual prior Completed label after saving a source edit remains KB-GUI-008; no product correction is claimed.","Initial capture-size and changed-button-caption helper assumptions were corrected and retained separately. They are not evidence that the product lacked a Run action.","Eighteen independent checks cover these stated routes. The prior Completed label after a saved source edit remains KB-GUI-008; stale-state acceptance is not passed. A Cancelled receipt retained an L1 row and is not accepted as a completed solve. Physical pickers, complete GUI article, final distributed package, other platforms, physical accuracy and mesh convergence remain pending. KB-GUI-006 single-tab restart workaround remains separate."],"scope":"Recorded inputs/builds only. Complete walkthrough and final-package acceptance remain separate.","verification_requirements":{"gui_method":"product-native scenario or recorded operator walkthrough; relevant screenshots required","status":"review-incomplete","illustrations":"required","required":["technical-review","release-applicability","execution","recovery","gui-workflow","illustrations"],"legacy_walkthrough_field":"pending retained; not a complete GUI article"},"sources":[{"id":"inductex:docs/manual/studio_projects.tex","section":"Raw IXI, saving, running and source/state identity"},{"id":"inductex:docs/manual/studio_results.tex","section":"Results, raw output, current and retained artifacts"},{"id":"validation:studio-inputs-native-windows-20261103","section":"VAL-023; exact source edit/Undo/Redo/Save, retained completion observation and successful new run; KB-GUI-008 remains open"}]},"title":"Results still says Completed after I edit an IXI file","type":"Troubleshooting","markdown_url":"/_functions/kbResource?path=markdown%2Finductex-completed-after-edit.md","summary":"The checked Studio candidate can retain the earlier Completed result after a saved source edit. Save the intended inputs, use Run again and check the new verdict and values.","product":"InductEx"}